Our personnel are always inside the spirit of continuous improvement and excellence, and together with the outstanding excellent goods, favorable price and good after-sales services, we try to gain every customer's trust for Micro Pressure Calibration Pump, Temperature Switch Calibration, Hydraulic High Pressure Calibration Pump, We welcome new and previous buyers from all walks of life to make contact with us for upcoming organization associations and mutual good results!
Hot sale Infrared Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran Detail:

Standard Platinum Resistance Thermometer

I. Description

The Standard Platinum Resistance Thermometer is used for compensation in standard temperature range of 13.8033k—961.8°C, and used as a standard when testing a variety of standard thermometers and high-precision thermometers. Within the above the temperature zone, it is also used directly for measuring the temperature of high-accuracy.

The Standard Platinum Resistance Thermometer measures the temperature according to the changeable regularity of the resistance temperature of the platinum.

In accordance with the regulations of ITS90, the T90 is defined by the platinum thermometer when triple point (13.8033K) of the nitrogen balance reaches to the temperature range of silver freezing point. It is indexed by using the group of required defined freezing point and reference function as well as the deviation function of the temperature interpolation.

The above temperature zoning is divided into several and decided to work normally within the sub-temperature zone by various different forms of structure of thermometers.

See the detailed thermometers in the table below:

Type

Classification

Suitable Temperature Zone

Working Length

(mm)

Temperature

WZPB-1

I

0~419.527

470±10

Medium

WZPB-1

I

83.8058K~419.527

470±10

Full

WZPB-2

II

0~419.527

470±10

Medium

WZPB-2

II

83.8058K~419.527

470±10

Full

WZPB-7

I

0~660.323

510±10

Medium

WZPB-7

I

83.8058K~660.323

510±10

Full

WZPB-8

II

0~660.323

510±10

Medium

WZPB-8

II

83.8058K~660.323

510±10

Full

Note: The Rtp of the above thermometers is 25±1.0Ω.The external diameter of quartz tubes is φ7±0.6mm.our factory also manufactures the platinum thermometer with the temperature zone of 83.8058K~660.323 as the working basic standard instrument.

II. Use Information

1. Before using, firstly, check the thermometer number to be consistent with the test certificate.

2. When using, according to the lug logo of the thermometer wire terminal, connect the wire correctly. The lug① of the red wire is connected to the current positive terminal; the lug of the yellow wire, to the current negative terminal; and the lug of the black wire, to the potential positive terminal; the lug of the green wire, to the potential negative terminal.

The following is the outline of the thermometer:

SPRT basic structure.png

3. The current should be 1MA according to the measuring of the temperature component of the thermometer.

4. For matching the electrical measurement device of the thermometer for measuring temperature, the low resistance potentiometer of grade 1 and the standard coil resistance of grade 0.1 or the measuring precise temperature bridge as well as the accessories should be used. The complete set of electrical measuring device should be guaranteed to have the sensitivity to distinguish the changing of one ten-thousandth Ohm.

5. During the process of using, preservation and transportation, try to avoid the severe mechanical vibration of the thermometer.

6. When using the first grade Standard Platinum Resistance Thermometer to test the temperature of the second grade Standard Platinum Resistance Thermometer, should follow the verification procedures approved by National Measurement Bureau.

7. The regular test of the thermometer should be done stringently according to the relevant verification procedures and regulations .


Product detail pictures:

Hot sale	Infrared Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Hot sale	Infrared Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Hot sale	Infrared Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Hot sale	Infrared Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Hot sale	Infrared Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures


Related Product Guide:

Our concentrate on should be to consolidate and enhance the quality and service of present products, meanwhile consistently produce new products to meet unique customers' demands for Hot sale Infrared Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran , The product will supply to all over the world, such as: United States, Croatia, Denmark, Our company will adhere to Quality first, , perfection forever, people-oriented , technology innovationbusiness philosophy. Hard work to keep making progress, innovation in the industry, make every effort to first-class enterprise. We try our best to build the scientific management model, to learn abundant skilled knowledge, to develop advanced production equipment and production process , to create the first-call quality solutions, reasonable price , high quality of service , quick delivery , to offer you create new value .
It can be said that this is a best producer we encountered in China in this industry, we feel lucky to work with so excellent manufacturer.
5 Stars By Alva from USA - 2017.12.31 14:53
The supplier cooperation attitude is very good, encountered various problems, always willing to cooperate with us, to us as the real God.
5 Stars By Sandy from Sierra Leone - 2017.01.28 19:59